Your Cart (0)
Thumbnail Filmstrip of P32 Precision Milliamp Current Loop Calibrator Images
P32 Precision Milliamp Current Loop Calibrator
Overview
The P32 is a handheld accurate process signal calibrator is adapted to the various process installations due to scaling function and specific ranges. It can simulate ramps, steps and fixed values to control process transmitters, panel meters and recorders. Temperature coefficients are low: 15 ppm /°C in voltage and 20 ppm/°C in current. The high accuracy of 150 ppm makes P32 process calibrator an ideal standard for onsite use even in demanding environmental conditions. General Features
Price Starting At:
$1,099.00
$1,099.00
Shipping Estimated Shipping 1 to 3 Weeks
Functions | Simulate and read milliamps, mA read/loop power and simulate and read volts |
Ranges and Accuracy | See Range and Accuracy Tables Below |
Resolution | 1uA on mA ranges and 1mV on voltage range | Operating Temperature | -10°C to 50°C (14 to 122°F) |
Operating Humidity Range | 20 to 80% non-condensing |
Reference Range | 23°C ±5°C (73 ±9°F) (RH: 45 to 75% non-condensing) |
Limit Operating Temperature | -15°C to 55°C (5 to 131°F), RH: 10 to 80% non-condensing (70% at 55°C) |
Storage Temperature | -30°C to 60°C (-22 to 140°F) |
Over-Voltage Protection | Electronic protection up to 250 V for ‘voltage’ wires' Fuse protection for ‘current’ wires Protection against ‘current’ circuit breaking during inductive resistance measurements |
Rated Voltage | 60 Volts |
Shocks and Vibrations | EN 61010-1 |
Display | 160 x 160 pixel liquid crystal graphical display with backlite Display of result as table of values or trend curve |
Power | AA alkaline or Optional rechargeable Ni-Mh batteries with internal charger in option |
Battery Life (using alkaline) | Approx. 53 hours in voltage and current measurement and voltage simulation Approx. 12 hours in current simulation |
EMC conformity | Immunity: EN 61000-4-2, EN 61000-4-3, EN 61000-4-5, EN 61000-4-6,EN 61000-4-11, EN 61000-4-4 Conducted and radiated emissions: EN 55022, class B, EN 61000-3-2, EN 61000-3-3 |
Size | 6.2" x 3.4" x 1.8" (157 x 85 x 45 mm) (H x W x D) |
Weight | 11 ounces (306 g) |
P32 Process Calibrator Simulation DC current With or without Loop Supply
Range | Simulation Range | Resolution | Accuracy/1yr |
---|---|---|---|
0 to 20 millamp | 500µA/24mA | 1 µA | 0.015% RDG + 2µA |
4 to 20 millamp | 3.2/24 mA | 1 µA | 0.015% RDG + 2 µA |
25 millamp | 500µA/24mA | 1 µA | 0.015% RDG + 2 µA |
Rising time: < 500 µs (0 to 20 mA accross 20Ω load)
VLF Noise: < 1 µA (at F < 100 Hz)
P32 Process Calibrator Measurement DC current With or without loop supply
Range | Simulation Range | Resolution | Accuracy/1yr |
---|---|---|---|
0 to 20 millamp | -6 to 24mA | 1 µA | 0.015% RDG +2µA |
4 to 20 millamp | 3.2/24 mA | 1 µA | 0.015% RDG + 2µA |
25 millamp | -6 to 25mA | 1 µA | 0.015% RDG + 2µA |
Loop power supply: 24 V ± 10%
HART® compatibility: R = 250Ω± 5%
Rin < 30Ω
Common mode rejection: 120 dB at 50 and 60 Hz
P32 Process Calibrator DC Voltage Simulation
Range | Simulation Range | Resolution | Accuracy/1yr |
---|---|---|---|
0 to 10 Volts | 0 to 12V | 1 mV | 0.015% RDG +2mV |
0 to 15 Volts | 0 to 15V | 1 mV | 0.015% RDG + 2mV |
Rising time: < 1 ms (0 V to 15 Volts accross 1 MΩ load)
Internal resistance: = 1Ω
VLF Noise: < 1 mVcc (at F < 100 Hz)
P32 Process Calibrator DC Voltage Measurement
Range | Simulation Range | Resolution | Accuracy/1yr |
---|---|---|---|
0 to 10 Volts | -2 to 12Volts | 1 mV | 0.015% RDG +2mV |
25 V | -2 to 25V | 1 mV | 0.015% RDG +2mV |
50 Volts | -5 to 50V | 1 mV | 0.015% RDG +4mV |
P32 Process Calibrator Continuity Test
Range | 0% | 25% | 25% | 75% | 100% |
---|---|---|---|---|---|
4 to 20 mA linear | 4 | 8 | 12 | 16 | 20 |
0 to 20 mA linear | 0 | 5 | 10 | 15 | 20 |
4 to 20 mA quad | 4 | 5 | 8 | 13 | 20 |