Secure Checkout
OPTCTLT15 - CT LT15 Compact Infrared Thermometer Sensor with LCD Display

Thumbnail Filmstrip of OPTCTLT15 - CT LT15 Compact Infrared Thermometer Sensor with LCD Display Images

    OPTCTLT15 - CT LT15 Compact Infrared Thermometer Sensor with LCD Display

    Overview

    The Optris Infrared Pyrometer OPTCTLT15 CT LT15 is equipped with one of the world’s smallest infrared sensors with a high optic distance to spot size resolution of 15:1. The sensor head is rugged and usable up to 356°F (180°C) ambient temperature without cooling. A separate electronics housing with easy accessible programming keys and LCD backlit display is included.
    Features
    • Temperature Range -58°F to 1112°F (-50°C to 600°C) scalable via programming keys or optional USB, RS232, interface and software
    • High Optic Resolution of 15:1
    • Spectral response 8 to 14µm
    • Compact Size M12x1, 1.1 in. (28mm) long, stainless steel sensor housing
    • Ambient temperature sensor head rugged and usable up to 356°F (180°C) ambient temperature without cooling
    • Separate electronics housing with easy accessible programming keys and LCD backlit display
    • IP 65 (NEMA-4) Sensing Head and Electronic Housing
    • Emissivity 0.100 to 1.000 adjustable via programming keys and LCD backlit display
    • Signal processing: Peak Hold, Valley Hold, Average; Extended hold function with Threshold and Hysteresis
    • Scalable analog output: 0/4 to 20mA, 0 to 5V, 0 to 10 V, or Thermocouple Type K or J
    • Optional USB, RS485, RS232 interface, relay outputs (2 x optically isolated)
    Price Starting At:
    $360.00
    Need help? Contact us with questions.

      Shipping Click Add to Cart to Select Options
    Please Call for Availability
    914-673-0333
    Optris OPTCTLT15 - CT LT15 IR Thermometer Measuring Specifications
    Temperature Range -58°F to 1112°F (-50°C to 600°C)
    Optical Resolution (Field of view)(90% energy) 15:1 (Precision Glass Optics)
    Spectral Range 8 to 14µm
    System Accuracy ±1.0% or ±1.8°F (±1.0°C) (at ambient temp. 73 ±9°F (23 ±5°C) Note 1, 2
    Repeatability ±0.5% or ±0.5°C (±0.9°F)(at ambient temp. 73 ±9°F (23 ±5°C)
    Temperature Resolution (Display) 0.1K
    NETD (Noise Equivalent Temperature Difference) 0.05K Note 2, 3
    Response Time (95%) 150 ms
    Emissivity 0.100 to 1.000 (adjustable via programming keys or optional USB, RS232, RS485 interface and software)
    Transmissivity 0.100 to 1.00 (adjustable via programming keys or optional USB, RS232, RS485 interface and software)
    Signal Processing Peak hold, valley hold, average; extended hold function with threshold and hysteresis.
    (Parameter adjustable via programming keys or software, respectively)
    OPTCTLT15 - CT LT15 IR Thermometer General Specifications
    Environmental Rating IP65 (NEMA-4) Sensor and Electronic Box
    Ambient Temperature Sensing head: -4°F to 356°F (-20°C to 180°C)
    Electronics: 32°F to 185°F (0°C to 85°C) Note 4
    Storage Temperature Sensing head: -40 °F to 266°F (-40°C to 130°C)
    Electronics: -40°F to 185°F (-40°C to 85°C)
    Relative Humidity 10 to 95%, non-condensing
    Dimensions Sensor: 1.1 in. (28mm) long X 0.55 in Dia.
    Electronic Housing: 3.62 in. (92mm) X 2.76 in.(70mm) X 1.18 in. (30mm)
    Vibration IEC 68-2-6: 3 G, 11 to 200Hz, any axis
    Shock IEC 68-2-27: 50G, 11ms, any axis
    Weight 40 g (1.4 oz) (sensing head)
    420 g (14.8 oz) (electronics)
    OPTCTLT15 CT LT15 IR Thermometer Electrical Specifications
    Output/analog Channel 1: 0/4 to 20mA, 0 to 5/10 V, Thermocouple Type J or K
    Channel 2: Sensing Head Temperature -4°F to 356°F (-20°C to 180°C) as 0 to 5V or 0 to 10V), alarm output
    Output/alarm 24V/50 mA (open collector)
    Optional Relay 2 x 60 V DC/42 V ACeff; 0.4 A; optically isolated
    Optional Output Digital USB, RS232, RS485, CAN-Bus, Profibus DP, Ethernet
    Output Impedances mA max. 500Ω (with 8-36 V DC)
    mV min. 100 kΩ load impedance
    thermocouple 20Ω
    Inputs Programmable functional input for external emissivity setting /ambient temperature compensation, trigger (reset of hold functions)
    Cable length 3.3ft (1m) standard. (Optional Lenghts 9.8ft (3m), 26.2ft (8m), 49.2ft (15m)
    Power Supply 8 to 36VDC
    Current Draw max. 100 mA
    Notes
    1) Whichever is greater
    2) At object temperatures > 0°C (32°F), Ɛ = 1
    3) At time constant 200 ms and TObj 25°C (77°F)
    4) The LCD displays capacity may be limited at ambient temperatures below 0°C (32°F).

    Each OPTCTLT15 CT LT15 includes Compact IR thermometer with stainless steel head (M12x1) mounting nut, and separate, configurable Electronic box with LCD display and programming keys, 1m cable and Operators manual,