Your Cart (0)
Thumbnail Filmstrip of C75 Documenting Multifunction Thermocouple, RTD Temperature and Process Signal Calibrator Images
C75 Documenting Multifunction Thermocouple, RTD Temperature and Process Signal Calibrator
Overview
The C75 is a field documenting multifunction temperature and process calibrator. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument.Features
- Simultaneously Measures and Simulates 14 Thermocouple Types, 12 RTD, Current, Resistance, Pressure, Voltage and Frequency
- DC current + Loop supply 24 V
- DC voltage 50V IN, 20V OUT
- Frequency 20KHz IN, 10KHz OUT
- Values displayed in °C, °F, Ω, Hz
- Pressure Measurement when used with an optional external pressure module, Relative pressure: 30 bar, Absolute pressure: 1.000 bar C75 can measure and simulate pressure (comparison calibration with a pressure pump)
- Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, synthetizer, square root, statistical functions/li>
- 10,000 data points stored and recalled on screen as curve or list
- Optional DATACAL Calibration Management Software Availabile
- IP54 according to EN60529
Price Starting At:
$5,205.00
$5,205.00
Shipping Estimated Shipping 1 to 4 Weeks
C75 Documenting Multifunction Temperature and Process Calibrator Specifications
Display | 240 x 320 pixel liquid crystal graphical display with backlite and contrast control |
Operating Temperature | -10 to 50°C (+14 to 122°F) |
Operating Humidity | 20 to 80% non-condensing |
Storage Temperature | -30°C to +60°C (-22°F to 140°F) |
Power Requirements | A pack of Ni-MH, 1.8 Ah rechargeable batteries |
Battery Life | 5 to 10 hours, depending on the functions used |
Dimensions | 8.3" W x 4.3" W x 2" D (210 x 110 x 50mm) |
Weight | 2 lbs. (900 g) |
CE - EMC | EN50082-1 and EN55022: 1994 Class B |
Memory | 10,000 data stored and recalled on screen as curve or list |
Optional Software | DATACAL calibration software for configuration and data management |
Serial Communications | USB |
Safety Protections | Electronic protection up to 250 V for ‘voltage’ wires Fuse protection for ‘current’ wires Protection against ‘current’ circuit breaking during inductive resistance measurements |
EMC conformity Immunity | EN 61000-4-2, EN 61000-4-3, EN 61000-4-5, EN 61000-4-6, EN 61000-4-11 |
Other | Waterproof to IP 54 in accordance with standard EN 60529 |
C75 Thermocouple Measurement and Simulation Range Specification Table
Download the Data Specification Sheet or the Users Manual for Complete Range and Accuracy Tables
Sensor Type | Temperature Input Range |
Temperature Input Resloution | Accuracy 1 Year (Measurement) | Temperature Output Range |
Temperature Output Resloution | Accuracy 1 Year (Simulation) |
---|---|---|---|---|---|---|
Type J | -210 to +1200°C (-346 to +2192°F) | 0.05°C | 0.013% RDG + 0.07°C(0.13°F) | -210 to +1200°C (-346 to +2192°F) | 0.05°C (0.09°F) | 0.013% RDG + 0.07°C(0.13°F) |
Type K | -250 to +1372°C (-418 to +2501°F) | 0.05°C (0.09°F) | 0.013% RDG + 0.08°C(0.14°F) | -240 to +1372°C (-400 to +2501°F)< | 0.2°C (0.36°F) | 0.013% RDG + 0.08°C(0.14°F) |
Type T | -250 to +400°C (-418 to +752°F) | 0.2°C (0.36°F) | 0.013% RDG + 0.08°C(0.144°F) | -240 to +400°C (-418 to +752°F) | 0.2°C (0.36°F) | 0.013% RDG + 0.08°C(0.144°F) |
Type E | -250 to +1000°C (-418 to +1832°F) | 0.1°C (0.18°F) | 0.013% RDG + 0.05°C(0.09°F) | -250 to +1000°C (-418 to +1832°F) | 0.1°C (0.18°F) | 0.013% RDG + 0.05°C(0.09°F) |
Type R | -50 to +1768°C (-58 to +3214°F) | 0.5°C (0.9°F) | 0.013% RDG + 0.2°C(0.36°F) | -50 to +1768°C (-58 to +3214°F) | 0.5°C (0.9°F) | 0.013% RDG +0.2°C(0.36°F) |
Type S | -50 to +1768°C (-58 to +3214°F) | 0.5°C (0.9°F) | 0.013% RDG + 0.25°C(0.45°F) | -50 to +1768°C (-58 to +3214°F) | 0.5°C (0.9°F) | 0.013% RDG +0.2°C(0.36°F) |
Type B | +400 to +1820°C (+752 to +3214°F) | 0.5°C (0.9°F) | 0.013% RDG + 0.25°C(0.45°F) | +400 to +1820°C (+752 to +3214°F) | 0.2°C (0.36°F) | 0.013% RDG +0.2°C(0.36°F) |
Type N | -240 to +1300°C (-400 to +2372°F) | 0.2°C (0.36°F) | 0.013% RDG + 0.15°C(0.108°F) | -240 to +1300°C (-400 to +2372°F) | 0.2°C (0.36°F) | 0.013% RDG + 0.06°C(0.108°F) |
Type C | -20 to +2310°C (+14 to +4190°F) | 0.1°C (0.18°F) | 0.013% RDG + 0.15°C(0.27°F) | -20 to +2310°C (+14 to +4190°F) | 0.1°C (0.18°F) | 0.013% RDG + 0.15°C(0.27°F) |
Type PLATINUM | -100 to +1400°C (-148 to +2552°F) | 0.05°C (0.18°F) | 0.3°C(0.54°F) | -100 to +1400°C (-148 to +2552°F) | 0.05°C (0.18°F) | 0.3°C(0.54°F) |
Type MO | 0 to +1375°C (+32 to +2507°F) | 0.05°C (0.18°F) | 0.013% RDG + 0.06°C(0.108°F) | 0 to +1375°C (+32 to +2507°F) | 0.05°C (0.18°F) | 0.013% RDG + 0.06°C(0.108°F) |
Type NiMo/NiCo | -50 to +1410°C (-58 to +2570°F) | 0.05°C (0.18°F) | 0.013% RDG + 0.30°C(0.54°F) | -50 to +1410°C (-58 to +2570°F)/td> | 0.05°C (0.18°F) | 0.013% RDG + 0.3°C(0.54°F) |
Type U | -200 to +660°C (-328 to +1220°F) | 0.05°C (0.09°F) | 0.15°C(0.27°F) | -200 to +600°C (-328 to +1112°F) | 0.05°C (0.09°F) | 0.15°C(0.27°F) |
Type L | -200 to +900°C (-328 to +1652°F) | 0.05°C (0.09°F) | 0.2°C(0.36°F) | -200 to +900°C (-328 to +1652°F) | 0.05°C (0.09°F) | 0.2°C(0.36°F) |
Accuracy is given for cold Junction reference @ 0°C (32°F)
When using the internal reference junction (except couple B) add an additional uncertainty of 0.2 °C at 0°C.
It is possible (thermocouple B excepted) to choose by programming the cold junction localization: External at 0°C, internal (temperature compensation of instrument’s terminals) or manually entered.
Temperature coefficient: <10% of accuracy/°C
Display unit: °C and °F
C75 RTD Measurement and Simulation Range Specification Table
Download the Data Specification Sheet or the Users Manual for Complete Range and Accuracy Tables
Sensor Type | Temperature Input Range | Temperature Input Resloution |
Accuracy 1 Year (Measurement) |
Accuracy 1 Year (Simulation) |
---|---|---|---|---|
Pt50 (α = 3851) | -220 to +850°C (-364 to +1562°F) | 0.01°C | 0.012% RDG + 0.06°C(0.108°F) | 0.014% RDG + 0.18°C(0.324°F) |
Pt100 (α = 3851) | -220 to +850°C (-364 to +1562°F) | 0.01°C | 0.012% RDG + 0.05°C(0.09°F) | 0.014% RDG + 0.12°C(0.216°F) |
Pt100 (α = 3916) | -200 to +510°C (-325 to +950°F) | 0.01°C | 0.012% RDG + 0.05°C(0.09°F) | 0.014% RDG + 0.12°C(0.216°F) |
Pt100 (α = 3926) | -210 to +850°C (-328 to +1562°F) | 0.01°C | 0.012% RDG + 0.05°C(0.09°F) | 0.014% RDG + 0.12°C(0.216°F) |
Pt200 (α = 3851) | -220 to +600°C (-364 to +1112°F) | 0.01°C | 0.012% RDG + 0.12°C(0.216°F) | 0.014% RDG + 0.33°C(0.594°F) |
Pt500 (α = 3851) | -220 to +600°C (-364 to +1112°F) | 0.01°C | 0.012% RDG + 0.07°C(0.126°F) | 0.014% RDG + 0.18°C(0.324°F) |
Pt1000 (α = 3851) | -60 to +850°C (-364 to +1112°F) | 0.01°C | 0.012% RDG + 0.05°C(0.126°F) | 0.014% RDG + 0.08°C(0.144°F) |
Ni100 (α = 618) | -60 to +180°C (-76 to +356°F) | 0.01°C | 0.012% RDG + 0.03°C(0.54°F) | 0.014% RDG + 0.08°C(0.144°F) |
Ni120 (α = 672) | -40 to +205°C (-40 to +401°F) | 0.01°C | 0.012% RDG + 0.03°C(0.54°F) | 0.014% RDG + 0.08°C(0.144°F) |
Ni1000 (α = 618) | -60 to +180°C (-76 to +356°F) | 0.01°C | 0.012% RDG + 0.03°C(0.54°F) | 0.014% RDG + 0.08°C(0.144°F) |
Cu10 (α = 427) | -50 to +150°C (-58 to +302°F) | 0.01°C | 0.012% RDG + 0.18°C(0.324°F) | 0.014% RDG + 0.1°C(0.18°F) |
Cu50 (α = 428 | -50 to +150°C (-58 to +302°F) | 0.01°C | 0.012% RDG + 0.06°C(0.108°F) | 0.014% RDG + 0.15°C(0.27°F) |
Accuracies are given for 4-wire mounted probes
Take into account particular error of temperature sensor used and implementation conditions
Measuring current: 0.25 mA (Measurement) or from 0.1 to 1 mA (Emission)
Temperature coefficient: < 10% of accuracy/°C
C75 Process Calibrator DC Current Simulation Table
With or without Loop SupplyRange | Resolution | Accuracy/1yr |
---|---|---|
24 millamp | 1µA | 0.0175% RDG + 2 µA |
0 to 20 millamp | 1µA | 0.0175% RDG + 2µA |
4 to 20 millamp | 1µA | 0.015% RDG + 2 µA |
Settling time: <5 ms
Specifications given for configurations in:
- Active mode (+24V ON) 1 Meter in passive mode (+24 V OFF)
- Passive mode (+24 V OFF) 1 Meter in active mode (+24 V ON)
C75 Process Calibrator DC Current Measurement Table
With or Without Loop SupplyRange | Resolution | Accuracy/1yr | Rin |
---|---|---|---|
0 to 20 millamp | 1µA | 0.0175% RDG + 2 µA | <25 Ω |
4 to 20 millamp | 1µA | 0.0175% RDG + 2 µA | <25 Ω |
±50 millamp | 1µA | 0.0175% RDG + 2 µA | <25 Ω |
Loop supply: 24V ±10%
HART® compatibility: Input impedance Rin = 280Ω
Display with linear or quadratic scaling
C75 Process Calibrator DC Voltage Simulation Table
Range | Resolution | Accuracy/1yr | Minimum Load |
---|---|---|---|
100 mV | 1 µV | 0.013% RDG + 3 µV | >1 kΩ |
2 Volts | 10 µV | 0.013% RDG + 20 µV | 2 kΩ |
20 Volts | 100 µV | 0.015% RDG +200 µV | 4 kΩ |
50 Volts | 1 mV | 0.015% RDG +2 mV | 4 kΩ |
Temperature coefficient: < 7 ppm/°C beyond reference domain
Settling time: < 5 ms
C55 Process Calibrator DC Voltage Measurement Table
Range | Resolution | Accuracy/1yr | Rin |
---|---|---|---|
100 mV | 1 µV | 0.013% RDG + 3 µV | >10 MΩ |
1 V* | 10 µV | 0.013% RDG + 20 µV | >10 MΩ |
10 Volts | 100 µV | 0.015% RDG + 200 µV | >1 MΩ |
50 Volts | 1 mV | 0.015% RDG + 200 Mv | >1 MΩ |
(^) Specification domain: -0.8 V to +1 V
C75 Process Calibrator Frequency, Pulse Simulation Table
Range | Resolution | Accuracy/1yr |
---|---|---|
1000 Hz | 0.01 Hz | 0.005% RDG |
10 kHz | 0.1 Hz | 0.005% RDG |
C75 Process Calibrator Frequency, Counting Measurement Table
Range | Resolution | Accuracy/1yr |
---|---|---|
20 kHz | 0.01 Hz | 0.005% RDG |
Scale unit: Pulse/min and Hz
Trigger level: 1 V
Measurement on frequency signals or dry contacts
Counting will be performed on defined time or infinite time.
C75 Process Calibrator Resistance Simulation Table
Range | Resolution | Accuracy/1yr | Notes |
---|---|---|---|
40 Ω | 1 mΩ | 0.014% RDG + 3 mΩ 0.014% RDG + 10 mΩ |
10 mA 1mA |
400 Ω | 10 mΩ | 0.014% RDG + 20 mΩ 0.014% RDG + 30 mΩ |
10 mA 1mA |
4000 Ω | 100 mΩ | 0.014% RDG + 300 mΩ | 0.1/1 mA |
Current establishing time: <1 ms
Notes: Current received by the calibrator
C75 Process Calibrator Resistance Measurement Table
Range | Resolution | Accuracy/1yr | Notes |
---|---|---|---|
400 Ω | 1 mΩ | 0.012% RDG + 10 mΩ | |
4000 Ω | 10 mΩ | 00.012% RDG + 100 mΩ |
Accuracies are given for 4-wire mounted probes
Temperature coefficient: <7 ppm/°C beyond reference domain
Open circuit terminal voltage: <10 V
Continuity test: Open circuit for R > 1000Ω and closed circuit for R <1000 Ω
Measurement current: 0.25mA
C75 Pressure: Measurement by external digital sensor Range Specification Table
Range | 0 to 1 bar | 0 to 3 bar | 0 to 10 bar | 0 to 30 bar | 0 to 100 bar | 0 to 300 bar | 0 to 1000 bar |
---|---|---|---|---|---|---|---|
Absolute | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
Relative | Yes | Yes | Yes | Yes |
Connector: ¼ gas
Resolution: 0.02% FS
Accuracy:-0.05% FS from 10 to 40°C; -0.1% FS from -10 to +10°C and from 40 to 80°C
This digital pressure module ACL433 is connected to C75 through RS485 serial cable to the digital input connector. All data are digital. Measurements are compensated in temperature by a polynomial correction implemented into the firmware at factory.
Each C75 Documenting Multifunction Temperature Process Calibrator includes User manual, Battery charger, Set of 6 testing leads and carrying strap.