Secure Checkout
C100 Precision Documenting Multifunction Temperature and Process Calibrator

Thumbnail Filmstrip of C100 Precision Documenting Multifunction Temperature and Process Calibrator Images

    C100 Precision Documenting Multifunction Temperature and Process Calibrator

    Overview

    The C100 is a high accuracy Field Documenting multifunction temperature and process calibrator. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument.
    Features
    • Simultaneously Measures and Simulates 14 Thermocouple Types, 12 RTD, Current, Resistance, Pressure, Voltage and Frequency
    • DC current + Loop supply 24 V
    • DC voltage 50V IN/20V OUT
    • Frequency 20KHz IN/10KHz OUT
    • Values displayed in °C, °F, Ω, Hz
    • Pressure Measurement when used with an optional external pressure module,
    • Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, synthetizer, square root, statistical functions/li>
    • C100 Integrates "Easyconnect" Terminals
    • 240 x 320 Pixel Liquid Crystal Graphical Display with Backlite
    • Display of minimum, maximum and average value
    • 10,000 data points stored and recalled on screen as curve or list
    • IP54 according to EN60529
    Price Starting At:
    $6,134.00
    Need help? Contact us with questions.

      Shipping Estimated Shipping 1 to 4 Weeks
    C100 Multifunction Temperature and Process Calibrator Specifications
    Display 240 x 320 pixel liquid crystal graphical display with backlite and contrast control
    Operating Temperature -10 to 50°C (+14 to 122°F)
    Operating Humidity 20 to 80% non-condensing
    Storage Temperature -30°C to +60°C (-22°F to 140°F)
    Power Requirements A pack of Ni-MH, 1.8 Ah rechargeable batteries
    Battery Life 5 to 10 hours, depending on the functions used
    Dimensions 8.3" W x 4.3" W x 2" D (210 x 110 x 50mm)
    Weight 2 lbs. (900 g)
    CE - EMC EN50082-1 and EN55022: 1994 Class B
    Memory 10,000 data stored and recalled on screen as curve or list
    Optional Software DATACAL calibration software for configuration and data management
    Serial Communications USB
    Safety Protections Electronic protection up to 250 V for ‘voltage’ wires
    Fuse protection for ‘current’ wires
    Protection against ‘current’ circuit breaking during inductive resistance measurements
    EMC conformity Immunity EN 61000-4-2, EN 61000-4-3, EN 61000-4-5, EN 61000-4-6, EN 61000-4-11
    Other Waterproof to IP 54 in accordance with standard EN 60529
    C100 Thermocouple Measurement and Simulation Range Specification Table
    Download the Data Specification Sheet or the Users Manual for Complete Range and Accuracy Tables
    Sensor Type Temperature Input Range Temperature Input Resloution Accuracy 1 Year (Measurement) Temperature Output Range Temperature Output Resloution Accuracy 1 Year (Simulation)
    Type J -210 to +1200°C (-346 to +2192°F) 0.05°C 0.010% RDG + 0.08°C(0.144°F) -210 to +1200°C (-346 to +2192°F) 0.05°C (0.09°F) 0.013% RDG + 0.07°C(0.126°F)
    Type K -250 to +1372°C (-418 to +2501°F) 0.05°C (0.09°F) 0.010% RDG + 0.08°C(0.14°F) -240 to +1372°C (-400 to +2501°F)< 0.2°C (0.36°F) 0.010% RDG + 0.08°C(0.14°F)
    Type T -250 to +400°C (-418 to +752°F) 0.2°C (0.36°F) 0.010% RDG + 0.08°C(0.144°F) -240 to +400°C (-418 to +752°F) 0.2°C (0.36°F) 0.010% RDG + 0.08°C(0.144°F)
    Type E -250 to +1000°C (-418 to +1832°F) 0.1°C (0.18°F) 0.010% RDG + 0.05°C(0.09°F) -250 to +1000°C (-418 to +1832°F) 0.1°C (0.18°F) 0.010% RDG + 0.05°C(0.09°F)
    Type R -50 to +1768°C (-58 to +3214°F) 0.5°C (0.9°F) 0.010% RDG + 0.2°C(0.36°F) -50 to +1768°C (-58 to +3214°F) 0.5°C (0.9°F) 0.010% RDG + 0.2°C(0.36°F)
    Type S -50 to +1768°C (-58 to +3214°F) 0.5°C (0.9°F) 0.003% RDG + 0.25°C(0.45°F) -50 to +1768°C (-58 to +3214°F) 0.5°C (0.9°F) 0.010% RDG +0.2°C(0.36°F)
    Type B +400 to +1820°C (+752 to +3214°F) 0.2°C (0.9°F) 0.010% RDG + 0.25°C(0.45°F) +400 to +1820°C (+752 to +3214°F) 0.5°C (0.36°F) 0.010% RDG +0.25°C(0.45°F)
    Type N -240 to +1300°C (-400 to +2372°F) 0.2°C (0.36°F) 0.010% RDG + 0.15°C(0.108°F) -240 to +1300°C (-400 to +2372°F) 0.2°C (0.36°F) 0.010% RDG + 0.06°C(0.108°F)
    Type C -20 to +2310°C (+14 to +4190°F) 0.1°C (0.18°F) 0.010% RDG + 0.15°C(0.27°F) -20 to +2310°C (+14 to +4190°F) 0.1°C (0.18°F) 0.010% RDG + 0.15°C(0.27°F)
    Type PLATINUM -100 to +1400°C (-148 to +2552°F) 0.05°C (0.18°F) 0.25°C(0.54°F) -100 to +1400°C (-148 to +2552°F) 0.05°C (0.18°F) 0.3°C(0.54°F)
    Type MO 0 to +1375°C (+32 to +2507°F) 0.05°C (0.18°F) 0.010% RDG + 0.06°C(0.108°F) 0 to +1375°C (+32 to +2507°F) 0.05°C (0.18°F) 0.010% RDG + 0.06°C(0.108°F)
    Type NiMo/NiCo -50 to +1410°C (-58 to +2570°F) 0.05°C (0.18°F) 0.010% RDG + 0.30°C(0.54°F) -50 to +1410°C (-58 to +2570°F)/td> 0.05°C (0.18°F) 0.010% RDG + 0.3°C(0.54°F)
    Type U -200 to +660°C (-328 to +1220°F) 0.05°C (0.09°F) 0.15°C(0.27°F) -200 to +600°C (-328 to +1112°F) 0.05°C (0.09°F) 0.15°C(0.27°F)
    Type L -200 to +900°C (-328 to +1652°F) 0.05°C (0.09°F) 0.2°C(0.36°F) -200 to +900°C (-328 to +1652°F) 0.05°C (0.09°F) 0.17°C(0.306°F)
    Thermocouples: D and G, for specifications, refer to the instruction manual (Please download) Accuracy is given for cold Junction reference @ 0°C (32°F)
    When using the internal reference junction (except couple B) add an additional uncertainty of 0.2 °C at 0°C.
    It is possible (thermocouple B excepted) to choose by programming the cold junction localization: External at 0°C, internal (temperature compensation of instrument’s terminals) or manually entered.
    Temperature coefficient: <10% of accuracy/°C
    Display unit: °C and °F
    C100 RTD Measurement and Simulation Range Specification Table
    Download the Data Specification Sheet or the Users Manual for Complete Range and Accuracy Tables
    Sensor Type Temperature Input Range Temperature Input Resloution Accuracy 1 Year (Measurement) Accuracy 1 Year (Simulation)
    Pt50 (a = 3851) -220 to +850°C (-364 to +1562°F) 0.01°C 0.01% RDG + 0.05°C(0.09°F) 0.012% RDG + 0.18°C(0.324°F)
    Pt100 (a = 3851) -220 to +850°C (-364 to +1562°F) 0.01°C 0.010% RDG + 0.05°C(0.09°F) 0.012% RDG + 0.12°C(0.216°F)
    Pt100 (a = 3916) -200 to +510°C (-325 to +950°F) 0.01°C 0.010% RDG + 0.05°C(0.09°F) 0.012% RDG + 0.12°C(0.216°F)
    Pt100 (a = 3926) -210 to +850°C (-328 to +1562°F) 0.01°C 0.010% RDG + 0.05°C(0.09°F) 0.012% RDG + 0.12°C(0.216°F)
    Pt200 (a = 3851) -220 to +600°C (-364 to +1112°F) 0.01°C 0.010% RDG + 0.12°C(0.216°F) 0.012% RDG + 0.33°C(0.594°F)
    Pt500 (a = 3851) -220 to +600°C (-364 to +1112°F) 0.01°C 0.010% RDG + 0.07°C(0.126°F) 0.012% RDG + 0.18°C(0.324°F)
    Pt1000 (a = 3851) -60 to +850°C (-364 to +1112°F) 0.01°C 0.010% RDG + 0.05°C(0.126°F) 0.012% RDG + 0.08°C(0.144°F)
    Ni100 (a = 618) -60 to +180°C (-76 to +356°F) 0.01°C 0.010% RDG + 0.03°C(0.54°F) 0.012% RDG + 0.08°C(0.144°F)
    Ni120 (a = 672) -40 to +205°C (-40 to +401°F) 0.01°C 0.010% RDG + 0.03°C(0.54°F) 0.012% RDG + 0.08°C(0.144°F)
    Ni1000 (a = 618) -60 to +180°C (-76 to +356°F) 0.01°C 0.010% RDG + 0.03°C(0.54°F) 0.012% RDG + 0.08°C(0.144°F)
    Cu10 (a = 427) -50 to +150°C (-58 to +302°F) 0.01°C 0.010% RDG + 0.18°C(0.324°F) 0.012% RDG + 0.1°C(0.18°F)
    Cu50 (a = 428 -50 to +150°C (-58 to +302°F) 0.01°C 0.010% RDG + 0.06°C(0.108°F) 0.012% RDG + 0.15°C(0.27°F)
    Resistive probes measurements in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen.
    Accuracies are given for 4-wire mounted probes
    Take into account particular error of temperature sensor used and implementation conditions
    Measuring current: 0.25 mA (Measurement) or from 0.1 to 1 mA (Emission)
    Temperature coefficient: < 10% of accuracy/°C
    C100 Process Calibrator DC Current Simulation Table
    With or without Loop Supply
    Range Resolution Accuracy/1yr
    24 millamp 1µA 0.012% RDG + 2 µA
    0 to 20 millamp 1µA 0.012% RDG + 2µA
    4 to 20 millamp 1µA 0.012% RDG + 2 µA
    Temperature Coefficient <10 ppm/°C beyond reference domain
    Settling time: <5 ms
    Specifications given for C100 configurations in:
    - Active mode (+24V ON) 1 Meter in passive mode (+24 V OFF)
    - Passive mode (+24 V OFF) 1 Meter in active mode (+24 V ON)
    C100 Process Calibrator DC Current Measurement Table
    With or Without Loop Supply
    Range Resolution Accuracy/1yr Rin
    0 to 20 millamp 1µA 0.012% RDG + 2 µA <25 Ω
    4 to 20 millamp 1µA 0.012% RDG + 2 µA <25 Ω
    ±50 millamp 1µA 0.012% RDG + 2 µA <25 Ω
    Temperature coefficient: <10 ppm/°C beyond reference domain
    Loop supply: 24V ±10%
    HART® compatibility: Input impedance Rin = 280Ω
    Display with linear or quadratic scaling
    C100 Process Calibrator DC Voltage Simulation Table
    Range Resolution Accuracy/1yr Minimum Load
    +100 mV 1 µV 0.010% RDG + 3 µV >1 kΩ
    +2 Volts 10 µV 0.010% RDG + 20 µV 2 kΩ
    +20 Volts 100 µV 0.012% RDG +200 µV<(1)/td> 4 kΩ
    +50 Volts 1 mV 0.012% RDG +2 mV 4 kΩ
    (1)Noise: 3 ppm (for 0.1 Hz to 10 Hz) and 5 ppm (for 10 Hz to 100 Hz)
    Temperature coefficient: < 7 ppm/°C beyond reference domain
    Settling time: < 5 ms
    C55 Process Calibrator DC Voltage Measurement Table
    Range Resolution Accuracy/1yr Rin
    +100 mV 1 µV 0.010% RDG + 3 µV >10 MΩ
    +1 V* 10 µV 0.010% RDG + 20 µV >10 MΩ
    +10 Volts 100 µV 0.010% RDG + 200 µV >1 MΩ
    +50 Volts 1 mV 0.015% RDG + 200 Mv >1 MΩ
    Temperature coefficient < 7 ppm/°C beyond reference domain
    (1) Range Specified: -0.8 V to +1 V
    C100 Process Calibrator Frequency, Pulse Simulation Table
    Range Resolution Accuracy/1yr
    1000 Hz 0.01 Hz 0.005% RDG
    10 kHz 0.1 Hz 0.005% RDG
    Temperature coefficient <5 ppm/°C beyond reference domain
    C100 Process Calibrator Frequency, Counting Measurement Table
    Range Resolution Accuracy/1yr
    20 kHz 0.01 Hz 0.005% RDG
    Temperature coefficient <5 ppm/°C beyond reference domain
    Scale unit: Pulse/min and Hz
    Trigger level: 1 V
    Measurement on frequency signals or dry contacts
    Counting will be performed on defined time or infinite time.
    C100 Process Calibrator Resistance Simulation Table
    Range Resolution Accuracy/1yr Notes
    40 Ω 1 mΩ 0.012% RDG + 3 mΩ
    0.012% RDG + 10 mΩ
    10 mA
    1mA
    400 Ω 10 mΩ 0.012% RDG + 20 mΩ
    0.012% RDG + 30 mΩ
    1/10 mA
    1mA
    4000 Ω 100 mΩ 0.012% RDG + 300 mΩ 0.1/1 mA
    Temperature coefficient: <5 ppm/°C beyond reference domain
    Current establishing time: <1 ms
    Notes: Current received by the calibrator
    C100 Process Calibrator Resistance Measurement Table
    Range Resolution Accuracy/1yr Notes
    400 Ω 1 mΩ 0.010% RDG + 10 mΩ
    4000 Ω 10 mΩ 00.010% RDG + 100 mΩ
    Resistance measurement in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
    Accuracies are given for 4-wire mounted probes
    Temperature coefficient: <7 ppm/°C beyond reference domain
    Open circuit terminal voltage: <10 V
    Continuity test: Open circuit for R > 1000Ω and closed circuit for R <1000 Ω
    Measurement current: 0.25mA
    C100 Pressure: Measurement by external digital sensor Range Specification Table
    Range 0 to 1 bar 0 to 3 bar 0 to 10 bar 0 to 30 bar 0 to 100 bar 0 to 300 bar 0 to 1000 bar
    Absolute Yes Yes Yes Yes Yes Yes Yes
    Relative Yes Yes Yes Yes
    Available in relative, absolute and differential pressure.
    Connector: ¼ gas
    Resolution: 0.02% FS
    Accuracy:-0.05% FS from 10 to 40°C; -0.1% FS from -10 to +10°C and from 40 to 80°C
    This digital pressure module ACL433 is connected to C100 through RS485 serial cable to the digital input connector. All data are digital. Measurements are compensated in temperature by a polynomial correction implemented into the firmware at factory.


    Each C100 Precission Documenting Multifunction Temperature Process Calibrator includes User manual, Battery charger, Set of 6 testing leads and carrying strap.