Secure Checkout
C50 Multifunction Temperature and Process Calibrator

Thumbnail Filmstrip of C50 Multifunction Temperature and Process Calibrator Images

    C50 Multifunction Temperature and Process Calibrator

    Overview

    The C50 Multifuncton Calibrator is the ideal tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process and frequency in one single instrument.
    Features
    • Simultaneously Measures and Simulates 14 Thermocouple Types, 12 RTD, Currecnt, Voltage and Frequency
    • Values displayed in °C, °F, Ω, Hz
    • Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, synthetizer, square root, statistical functions
    • C50 Integrates "Easyconnect" Terminals
    • 240 x 320 Pixel Liquid Crystal Graphical Display with Backlite
    • Display of minimum, maximum and average value
    • Simple and cyclical ramp generation
    • IP54 according to EN60529
    Price Starting At:
    $2,799.00
    Need help? Contact us with questions.

      Shipping Estimated Shipping 1 to 4 Weeks
    C50 Multifunction Temperature and Process Calibrator Specifications
    Display 240 x 320 pixel liquid crystal graphical display with backlite and contrast control
    Operating Temperature -10 to 50°C (+14 to 122°F)
    Operating Humidity 20 to 80% non-condensing
    Storage Temperature -30°C to +60°C (-22°F to 140°F)
    Power Requirements Ni-MH, 1.8 Ah rechargeable batteries
    Battery Life 5 to 10 hours, depending on the functions used
    CE - EMC EN50082-1 and EN55022: 1994 Class B
    Serial Communications USB
    Safety Protections Electronic protection up to 250 V for ‘voltage’ wires
    Fuse protection for ‘current’ wires
    Protection against ‘current’ circuit breaking during inductive resistance measurements
    EMC conformity Immunity EN 61000-4-2, EN 61000-4-3, EN 61000-4-5, EN 61000-4-6, EN 61000-4-11
    Dimensions 8.3" W x 4.3" W x 2" D (210 x 110 x 50mm)
    Weight 2 lbs. (900 g)
    Other Waterproof to IP 54 in accordance with standard EN 60529
    C50 Thermocouple Measurement and Simulation Range Specification Table
    Download the Data Specification Sheet or the Users Manual for Complete Range and Accuracy Tables
    Sensor Type Temperature Input Range Temperature Input Resloution Accuracy 1 Year (Measurement) Temperature Output Range Temperature Output Resloution Accuracy 1 Year (Simulation)
    Type J -210 to +1200°C (-346 to +2192°F) 0.05°C 0.013% RDG + 0.07°C(0.13°F) -210 to +1200°C (-346 to +2192°F) 0.05°C (0.09°F) 0.013% RDG + 0.07°C(0.13°F)
    Type K -250 to +1372°C (-418 to +2501°F) 0.05°C (0.09°F) 0.013% RDG + 0.08°C(0.14°F) -240 to +1372°C (-400 to +2501°F) 0.2°C (0.36°F) 0.013% RDG + 0.08°C(0.14°F)
    Type T -250 to +400°C (-418 to +752°F) 0.2°C (0.36°F) 0.013% RDG + 0.08°C(0.144°F) -240 to +400°C (-418 to +752°F) 0.2°C (0.36°F) 0.013% RDG + 0.08°C(0.144°F)
    Type E -250 to +1000°C (-418 to +1832°F) 0.1°C (0.18°F) 0.013% RDG + 0.05°C(0.09°F) -250 to +1000°C (-418 to +1832°F) 0.1°C (0.18°F) 0.013% RDG + 0.05°C(0.09°F)
    Type R -50 to +1768°C (-58 to +3214°F) 0.5°C (0.9°F) 0.013% RDG + 0.2°C(0.36°F) -50 to +1768°C (-58 to +3214°F) 0.5°C (0.9°F) 0.013% RDG +0.2°C(0.36°F)
    Type S -50 to +1768°C (-58 to +3214°F) 0.5°C (0.9°F) 0.013% RDG + 0.25°C(0.45°F) -50 to +1768°C (-58 to +3214°F) 0.5°C (0.9°F) 0.013% RDG +0.25°C(0.45°F)
    Type B +400 to +1820°C (+752 to +3214°F) 0.5°C (0.9°F) 0.013% RDG + 0.25°C(0.45°F) +400 to +1820°C (+752 to +3214°F) 0.2°C (0.36°F) 0.013% RDG +0.2°C(0.36°F)
    Type N -240 to +1300°C (-400 to +2372°F) 0.2°C (0.36°F) 0.013% RDG + 0.15°C(0.108°F) -240 to +1300°C (-400 to +2372°F) 0.2°C (0.36°F) 0.013% RDG + 0.06°C(0.108°F)
    Type C -20 to +2310°C (+14 to +4190°F) 0.1°C (0.18°F) 0.013% RDG + 0.15°C(0.27°F) -20 to +2310°C (+14 to +4190°F) 0.1°C (0.18°F) 0.013% RDG + 0.15°C(0.27°F)
    Type PLATINUM -100 to +1400°C (-148 to +2552°F) 0.05°C (0.18°F) 0.3°C(0.54°F) -100 to +1400°C (-148 to +2552°F) 0.05°C (0.18°F) 0.3°C(0.54°F)
    Type MO 0 to +1375°C (+32 to +2507°F) 0.05°C (0.18°F) 0.013% RDG + 0.06°C(0.108°F) 0 to +1375°C (+32 to +2507°F) 0.05°C (0.18°F) 0.013% RDG + 0.06°C(0.108°F)
    Type NiMo/NiCo -50 to +1410°C (-58 to +2570°F) 0.05°C (0.18°F) 0.013% RDG + 0.30°C(0.54°F) -50 to +1410°C (-58 to +2570°F)/td> 0.05°C (0.18°F) 0.013% RDG + 0.30°C(0.54°F)
    Type U -200 to +660°C (-328 to +1220°F) 0.05°C (0.09°F) 0.15°C(0.27°F) -200 to +600°C (-328 to +1112°F) 0.05°C (0.09°F) 0.15°C(0.27°F)
    Type L -200 to +900°C (-328 to +1652°F) 0.05°C (0.09°F) 0.2°C(0.36°F) -200 to +900°C (-328 to +1652°F) 0.05°C (0.09°F) 0.2°C(0.36°F)
    Thermocouples: D and G, for specifications, refer to the instruction manual (Please download) Accuracy is given for cold Junction reference @ 0°C.
    When using the internal reference junction (except couple B) add an additional uncertainty of 0.2 °C at 0°C.
    It is possible (thermocouple B excepted) to choose by programming the cold junction localization: External at 0°C, internal (temperature compensation of instrument’s terminals) or manually entered.
    Temperature coefficient: <10% of accuracy/°C
    Display unit: °C and °F
    C50 RTD Measurement and Simulation Range Specification Table
    Download the Data Specification Sheet or the Users Manual for Complete Range and Accuracy Tables
    Sensor Type Temperature Input Range Temperature Input Resloution Accuracy 1 Year (Measurement) Accuracy 1 Year (Simulation)
    Pt50 (α = 3851) -220 to +850°C (-364 to +1562°F) 0.01°C 0.012% RDG + 0.06°C(0.108°F) 0.014% RDG + 0.18°C(0.324°F)
    Pt100 (α = 3851) -220 to +850°C (-364 to +1562°F) 0.01°C 0.012% RDG + 0.05°C(0.09°F) 0.014% RDG + 0.12°C(0.216°F)
    Pt100 (α = 3916) -200 to +510°C (-325 to +950°F) 0.01°C 0.012% RDG + 0.05°C(0.09°F) 0.014% RDG + 0.12°C(0.216°F)
    Pt100 (α = 3926) -210 to +850°C (-328 to +1562°F) 0.01°C 0.012% RDG + 0.05°C(0.09°F) 0.014% RDG + 0.12°C(0.216°F)
    Pt200 (α = 3851) -220 to +600°C (-364 to +1112°F) 0.01°C 0.012% RDG + 0.12°C(0.216°F) 0.014% RDG + 0.33°C(0.594°F)
    Pt500 (α = 3851) -220 to +600°C (-364 to +1112°F) 0.01°C 0.012% RDG + 0.07°C(0.126°F) 0.014% RDG + 0.18°C(0.324°F)
    Pt1000 (α = 3851) -60 to +850°C (-364 to +1112°F) 0.01°C 0.012% RDG + 0.05°C(0.126°F) 0.014% RDG + 0.08°C(0.144°F)
    Ni100 (α = 618) -60 to +180°C (-76 to +356°F) 0.01°C 0.012% RDG + 0.03°C(0.54°F) 0.014% RDG + 0.08°C(0.144°F)
    Ni120 (α = 672) -40 to +205°C (-40 to +401°F) 0.01°C 0.012% RDG + 0.03°C(0.54°F) 0.014% RDG + 0.08°C(0.144°F)
    Ni1000 (α = 618) -60 to +180°C (-76 to +356°F) 0.01°C 0.012% RDG + 0.03°C(0.54°F) 0.014% RDG + 0.08°C(0.144°F)
    Cu10 (α = 427) -50 to +150°C (-58 to +302°F) 0.01°C 0.012% RDG + 0.18°C(0.324°F) 0.014% RDG + 0.1°C(0.18°F)
    Cu50 (α = 428 -50 to +150°C (-58 to +302°F) 0.01°C 0.012% RDG + 0.06°C(0.108°F) 0.014% RDG + 0.15°C(0.27°F)
    Resistive probes measurements in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen.
    Accuracies are given for 4-wire mounted probes
    Take into account particular error of temperature sensor used and implementation conditions
    Measuring current: 0.25 mA (Measurement) or from 0.1 to 1 mA (Emission)
    Temperature coefficient: < 10% of accuracy/°C
    C50 Process Calibrator Simulation DC Current
    With or without Loop Supply
    Range Resolution Accuracy/1yr
    24 millamp 1µA 0.0175% RDG + 2 µA
    0 to 20 millamp 1µA 0.0175% RDG + 2µA
    4 to 20 millamp 1µA 0.0175% RDG + 2 µA
    Temperature Coefficient <10 ppm/°C beyond reference domain
    Settling time: <5 ms
    Specifications given for configurations in:
    - Active mode (+24V ON) 1 Meter in passive mode (+24 V OFF)
    - Passive mode (+24 V OFF) 1 Meter in active mode (+24 V ON)
    C50 Process Calibrator Measurement DC Current
    With or Without Loop Supply
    Range Resolution Accuracy/1yr Rin
    0 to 20 millamp 1µA 0.0175% RDG + 2 µA <25 Ω
    4 to 20 millamp 1µA 0.0175% RDG + 2 µA <25 Ω
    ±50 millamp 1µA 0.0175% RDG + 2 µA <25 Ω
    Temperature coefficient: <10 ppm/°C beyond reference domain
    Loop supply: 24V ±10%
    HART® compatibility: Input impedance Rin = 280Ω
    Display with linear or quadratic scaling
    C50 Process Calibrator Simulation DC Voltage
    Range Resolution Accuracy/1yr Minimum Load
    100 mV 1 µV 0.013% RDG + 3 µV 1 kΩ
    2 Volts 10 µV 0.015% RDG + 20 µV 2 kΩ
    20 Volts 100 µV 0.013% RDG +200 µV 4 kΩ
    Noise: 3 ppm (for 0.1 Hz to 10 Hz) and 5 ppm (for 10 Hz to 100 Hz)
    Temperature coefficient: < 7 ppm/°C beyond reference domain
    Settling time: < 5 ms
    C50 Process Calibrator Measurement DC Voltage
    Range Resolution Accuracy/1yr Rin
    100 mV 1 µV 0.013% RDG + 3 µV >10 MΩ
    1 V* 10 µV 0.013% RDG + 20 µV >10 MΩ
    10 Volts 100 µV 0.015% RDG + 200 µV >1 MΩ
    50 Volts 1 mV 0.015% RDG + 2mV >1 MΩ
    Temperature coefficient < 7 ppm/°C beyond reference domain
    (^) Specification domain: -0.8 V to +1 V
    C50 Process Calibrator Simulation Frequency Pluse
    Range Resolution Accuracy/1yr
    1000 Hz 0.01 Hz 0.005% RDG
    10 kHz 0.1 Hz 0.005% RDG
    Temperature coefficient <5 ppm/°C beyond reference domain
    C50 Process Calibrator Measurement Frequency, Counting
    Range Resolution Accuracy/1yr
    20 kHz 0.01 Hz 0.005% RDG
    C50 Process Calibrator Simulation Resistance
    Range Resolution Accuracy/1yr Notes
    40 Ω 1 mΩ 0.014% RDG + 3 mΩ
    0.014% RDG + 10 mΩ
    10 mA
    1mA
    400 Ω 10 mΩ 0.014% RDG + 20 mΩ
    0.014% RDG + 30 mΩ
    10 mA
    1mA
    4000 Ω 100 mΩ 0.014% RDG + 300 mΩ 0.1/1 mA mA
    1mA
    Temperature coefficient: <5 ppm/°C beyond reference domain
    Current establishing time: < 1 ms
    Notes: Current received by the calibrator
    C50 Process Calibrator Resistance Measurement
    Range Resolution Accuracy/1yr Notes
    400 Ω 1 mΩ 0.012% RDG + 10 mΩ 10 mA
    1mA
    4000 Ω 10 mΩ 00.012% RDG + 100 mΩ 10 mA
    1mA
    Resistance measurement in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
    Accuracies are given for 4-wire mounted probes
    Temperature coefficient: < 7 ppm/°C beyond reference domain
    Open circuit terminal voltage: < 10 V
    Continuity test: Open circuit for R > 1000 O and closed circuit for R < 1000 Ω
    Measurement current: 0.25 mA


    Eash C50 Multifunction Temperature Process Calibrator includes User manual, Battery charger, Set of 6 testing leads and carrying strap.